国产日韩欧美一区二区东京热,一本精品99久久精品77,亚洲精品欧美综合四区,国产亚洲精久久久久久无码77777

EMAIL: sales@hplc.com.cn

LANGUAGE:

Science Technology (Hangzhou) Inc.

The Current Qosition:Home |Products  | HPLC

LC-10TIsocratic HPLC

Share:

Product Introduction

uick Details Flow: 0.001 – 9.999 ml/min Max Pressure: 6000 psi Number of Solvent: 1 Minimum detection limit: 1×10-9g/ml Brand Name: Surwit Model: LC-10Tvp Certificate:GB/T 19001-2008/ISO 9001:2008

LC-10Tvp Isocratic Standard Configuration
 
LC-10Tvp Liquid Micro volume series double plunger
1 set
SPD-10Tvp UV-VIS Spectrophotometric Detector
1 set
SCL-10Tvp System Controller
1 set
7725i manual injector
1 set
Chromatography data system
1 set
Chromatography column
1 set


The LC-10Tvp Liquid Chromatograph System is designed by the Research and development team of Surwit Technology Inc.. It can be widely used in the field of medicine, food, chemical industry and environment.
 
In either case, the system incorporates several key features which make it truly one of a kind in the world of HPLC Systems. The LC-10 liquid system is more durable and stable as all the key parts are imported from USA.
 

Trinity design

LC-10 Tvp adopted the trinity system of column over, system controller and solvent tray as a whole, which can reasonably used the space and enhanced the management of temperature and solvent.
 
Audit trail and e-signature
Conform to GLP/GMP、FDA、21CFR Part Ⅱ standard.
 
Intuitive Software Control
The LC-10Tvp Liquid Chromatograph System adopts the most advanced liquid chromatography system called VI2010 HPLC Data WorkStation, which designed for international standard, all the action of LC-10T can be directly controlled by the computer. The control is designed with the dedicated keys that allow even a novice to turn on the pump and create methods. An easy to read data display allows the user to monitor several parameters from the same screen. When configured as a gradient HPLC system user can program and store different methods, all the methods can be linked or run independently.
 
Easy maintenance
All user serviceable components are easily accessed from the front panel. Once the head assembly is removed, the pistons and seals can be replaced quickly and easily without the use of tools. The highest voltage of LC-10Tvp can be reached to 10000 psi, and the minimum detection limit is 1×10-9g/ml.


Features and Specification

 

LC-10Tvp Liquid Pump

 

Infusion way

Series dual-plunger pump

Pressure Accuracy

± 2%

Flow Rates

0.001 ~9.999 ml/min

Flow Accuracy

≤0.5%

Pressure

0 - 9999 psi

Flow Precision

≤0.2% RSD

 

Pulsation

<15psi from 600 to 1,600 psi, Flow   1mL/min

Operating Temperature

4~40

Dimensions

130H x 260W x 420 D, (mm)

Weight.

11 kg

 

SPD-10Tvp UV-VIS Spectrophotometric Detector

 

Wavelength

190-700 nm

Spectral bandwidth

6 nm

Wavelength Accuracy

≤1 nm

Flow pool volume

8 μ L

Wavelength Reproducibility

≤± 0.1 nm

Optical path

10 mm

Dynamic Noise

≤± 0.75×10-5AU (Methanol, 1ml/min, 254nm,   20°C)

Static Noise

≤± 0.5×10-5AU(Empty pool, Response time 1 s,   20 °C)

Dynamic Drifting

≤± 1×10-4AU (Methanol, 1ml/min, 254nm, 20°C)

Static Drifting

≤± 0.5×10-4AU(Empty pool, Response time 1 s,   20 °C)

Qualitative repeatability

RSD6 ≤ 0.1%

Quantitative repeatability

RSD6 ≤ 0.5%

Dimensions

130H x 260W x 420 D, (mm)

Weight.

(11 kg)

Operating Temperature

4~40°C